Quiz

Right Atrial Autothreshold (RAAT)

Right Atrial Autothreshold (RAAT)

The device selects one of the following 2 methods, P test and AR test, to evaluate atrial capture based on the patient’s rhythm at the time of the pacing threshold search

  • RAAT is launched every night at 1:00 am
  • a selection phase is required to identify which test will be used
  • P test runs during normal stable sinus rhythm; atrial threshold is measured by increasing the pacing voltage
  • AR test runs when stable 1:1 AV conduction is observed with atrial pacing; atrial threshold is measured by decreasing the pacing voltage

 

Selection phase

 

P test

The P test is launched if the device has detected 6 PP intervals (7 P waves) over the last 8 cycles and the PP intervals are stable (<16 ms).

  • overdriving of the atrial rate
  • progressive increase of the atrial pacing amplitude (steps of 0.25 V)
  • the diagnosis of atrial capture/non capture is based on the absence/presence of a sensed atrial signal following atrial pacing
  • if there is one As after Ap: no atrial capture
  • if there is no As after Ap: atrial capture

 

 

In order to check if the atrial spike is efficient (capture), the device calculates the PP stability window (PP average +/- 8 ms).

For each tested amplitude:

  • the device senses 3 atrial signals that have to be within the stability window
  • it delivers atrial pacing (spontaneous atrial rate – 200 ms to overdrive sinus rhythm)
  • if an atrial signal is detected within the stability window: diagnosis of non capture and increase of the amplitude by 0.25 V steps
  • if no atrial signal is detected within the stability window: diagnosis of capture; if a second capture occurs at the same amplitude, the test value corresponds to the atrial threshold

 

AR test

The device verifies the presence of a stable 1:1 AV conduction by prolonging the AV delay (450 ms) during 10 cycles (2 first cycles of transition).

  • progressive decrease in the atrial pacing amplitude
  • the diagnosis of atrial capture/non capture is based on the absence/presence of a sensed ventricular signal following atrial pacing
  • if there is one R wave (Vs) after Ap: capture
  • if there is no R wave (Vs) after Ap: no capture (atrial back-up pacing)

 

In order to check if the atrial spike is efficient (capture), the device calculates the Ap-Vs stability window (Ap-Vs average +/- 50 ms).

For each tested amplitude:

  • decrease of the amplitude by 0.25 V steps
  • the device delivers 3 control spikes (at 2 V or 5 V) + 1 test spike (tested amplitude)
  • if a ventricular signal is detected within the stability window: capture
  • if a ventricular signal is not detected within the stability window: non capture; atrial backup pacing at the end of the stability window; second test at the same amplitude; if non efficient: atrial threshold previous tested value

 

Adjustment of the amplitude

  • the minimum atrial amplitude is programmable
  • automatic adjustment of the atrial pacing amplitude: measured atrial threshold + 1 V
  • if the measured threshold is > 3 V, the amplitude is automatically programmed to 5 V and RAAT is switched to OFF
  • if RAAT cannot be measured, the device takes the same value as the day before
  • if RAAT cannot be measured during more than 7 consecutive days, the atrial amplitude is programmed to the Safety atrial amplitude (programmable, nominal 3.5 V)

 

Programmable parameters

  • RAAT: Auto, “Monitor”, Off
  • minimum atrial amplitude: 1.0 – 5 – 2.0 – 2.5 V
  • safety atrial amplitude: 2.5 – 5 – 4.0 – 5.0 V
  • atrial autothreshold max rate: 75 – 80 – 85 – 90 – 95 – 100 – 110 min-1
Share this page: